Ausstellung in Ontario, Kalifornien · Support am Zielort anfragen
Startseite / Maschinen / Leckprüfung / Sealing Tester / Leak Tester
SED-MCY-A
Auf Anfrage
SED-MCY-A Sealing Tester / Leak Tester
SED · SED-MCY-A · 9 verkaufte Einheiten

SED-MCY-A — Sealing Tester / Leak Tester

Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English

SED-MCY-A is a vacuum sealing tester (leak tester) designed and manufactured per GB/T 15171-94 — test method for sealing performance of flexible packaging — widely used for QC of food, pharmaceutical, daily-chemical and

Variante wählen (2)
Lieferzeit
Auf Anfrage
Durchsatz
12–60 L/min extraction rate (vacuum-pump version)

Funktionsweise

Sample is placed inside the acrylic transparent vacuum tank with water around it (water level ~10 mm above the perforated press plate). The instrument evacuates the tank via either a compressed-air-driven Venturi vacuum generator (compressor version) or an external rotary-vane vacuum pump (vacuum-pump version). When the tank pressure reaches the set negative pressure ±accuracy band, the pump cycles off; if pressure rises above the band, the pump kicks back in. Sealed packaging in the water has higher internal pressure than the surrounding tank — leaks emit a stream of bubbles visible through the acrylic. After the user-set hold time elapses (1–9999 s), the instrument automatically opens the relief solenoid valve, returning tank pressure to atmospheric.

Prozessbeschreibung
1Sample is placed inside the acrylic transparent vacuum tank with water around it (water level ~10 mm above the perforated press plate).
2The instrument evacuates the tank via either a compressed-air-driven Venturi vacuum generator (compressor version) or an external rotary-vane vacuum pump (vacuum-pump version).
3When the tank pressure reaches the set negative pressure ±accuracy band, the pump cycles off; if pressure rises above the band, the pump kicks back in.
4Sealed packaging in the water has higher internal pressure than the surrounding tank — leaks emit a stream of bubbles visible through the acrylic.
5After the user-set hold time elapses (1–9999 s), the instrument automatically opens the relief solenoid valve, returning tank pressure to atmospheric.

Technische Daten

Technische Daten für Variante: SED-MCY-A air-compressor version (Venturi generator

Technische Daten

Modell
SED-MCY-A
Durchsatz
12–60 L/min extraction rate (vacuum-pump version)
Spannung
AC 220V 50/60Hz, 200 W
Versorgung
AC 220V 50/60Hz, 200 W. Compressed air at 0.7 MPa for compressor version OR external rotary-vane vacuum pump (VALUE FY-1H-N) for vacuum-pump version. Reliable grounding.
Konformitätsangaben
Designed per GB/T 15171-94 for flexible packaging sealing test. Reliable grounding required. Cannot be used in high-humidity, strong-corrosive-gas, strong-vibration or high-EMI environments.
Leistung
0.2 kW
Luftdruck
Pressure source 0.7 MPa (7 kgf/cm2) for compressor variant; vacuum 0-90 kPa (compressor) or 0-95 kPa (vacuum-pump)

Abmessungen

Länge
300 mm
Breite
380 mm
Höhe
200 mm
Gewicht
18 kg
Aufstellfläche
Controller 300 × 380 × 200 mm benchtop; vacuum tank Φ270 × 210 mm H standard, optional Φ360 × 585 mm or Φ460 × 330 mm. Reserve ≥1 m clearance. Ambient 10–40 °C; clean, dry, low-EMI environment.

Werkstoffe

Kontaktwerkstoffe
Acrylic transparent vacuum tank; SMC pneumatic fittings; PU 6 mm and 8 mm air tubing

Betrieb

Operating Humidity
<=85% RH
Operating Temperature
5-35 degC °C
Vacuum Accuracy
class 0.5
Vacuum Hold Time
2.5h h
Vacuum Range
0 to -90 kPa kPa
Resolution
0.1 kPa kPa

Physikalische Angaben

Air Port Size
Ø6mm polyurethane tube mm
Chamber Size
Ø270mm x 210mm H (standard)

Wesentliche Merkmale

Vacuum range 0–90 kPa (compressor) or 0–95 kPa (vacuum-pump variant)
Pressure-display accuracy ±0.5 kPa, resolution 0.1 kPa
Vacuum hold time 1–9999 seconds (~277 hours)
Maximum holding pressure −99.9 kPa
Settable accuracy / hysteresis band per test recipe
Extraction rate 12 L/min (compressor) or 60 L/min (vacuum pump)
Acrylic transparent vacuum tank for visual leak observation
Standard tank Φ270 × 210 mm H; optional Φ360 × 585 mm and Φ460 × 330 mm
LCD display + touch-button human-machine interface (Chinese/English)
Auto pressurise / hold / release cycle with intelligent control
Manual mode (Start / Stop / Blow Back) for diagnostic operation
Designed to GB/T 15171-94
Pressure calibration function with on-screen wizard (password 3052)
SMC pneumatic components (Japan) — reliable, low maintenance
Front-panel result-record button for pass/fail logging
Net weight 18 kg; controller 300 × 380 × 200 mm; AC 220V / 200 W
Auto-detect zero point and altitude-compensation calibration
Beep indicator for last 5 seconds of countdown
Two vacuum-source options — reuse existing air compressor or attach VALUE rotary-vane pump
Acrylic transparent tank lets QC visually confirm leaks — simple training
Designed to national standard GB/T 15171-94 — export-ready
Wide tank size options for ampoules, blister, pouches, large bottles
SMC pneumatics + electronic pressure sensor — lab-grade accuracy
Lifetime maintenance support after 12-month warranty
Compact bench size suits any QC lab

Anwendungen

Sealing test of flexible food / pharma / daily-chemical packaging
Vacuum leak test of blister packs, sachets, foil pouches
Sealing test of ampoules, vials, sealed bottles
Confirmation of packaging integrity after drop / pressure tests
QC sampling on production lines or incoming inspection

Branchen

Pharmaceutical and nutraceutical packaging QC
Food and beverage packaging
Cosmetics / personal care
Daily chemical
Medical devices / sterile pouches
Laboratory / R&D

Komponentenhersteller

SMC, Japan (core pneumatic components)
VALUE/(FY-1H-N 1 L mini rotary-vane vacuum pump,vacuum-pump variant)
Acrylic vacuum tank (SED in-house)
SED
Sprechen Sie mit SED Sales Team
Maschinenspezialisten · Kontakt per WhatsApp
Geschäftszeiten und Verfügbarkeit beim Team erfragen

Reinigung und Wartung

Keep air source dry and clean; regularly clean inlet filter. Drain water from air compressor regularly. Keep equipment and surroundings clean. Periodically check pressure-display accuracy; recalibrate if drift is found (calibration password 3052). Check vacuum-pump oil level — keep between upper and lower marks; replace oil when emulsified or contaminated.

Dokumentation

Datenblatt (PDF)
Überblick, technische Daten und Abmessungen · Originaldokument (Englisch)
Beim Vertrieb anfragen
Bedienhandbuch
Einrichtung, Betrieb und Fehlersuche · Originaldokument (Englisch)
Beim Vertrieb anfragen
Vergleichsdatenblatt (Standard und Economy)
Vergleich der Konfigurationen · Originaldokument (Englisch)
Beim Vertrieb anfragen
IQ / OQ / PQ-Protokolle
Validierungsunterlagen zur GMP-Konformität · Originaldokument (Englisch)
Beim Vertrieb anfragen
FAT / SAT-Protokolle
Abnahmeprüfungen im Werk und vor Ort · Originaldokument (Englisch)
Beim Vertrieb anfragen
CAD-Zeichnungen (STEP / IGES / DWG)
Konstruktionszeichnungen für die Aufstellung · Originaldokument (Englisch)
Beim Vertrieb anfragen
Kalibrierzertifikate
Schallpegel, Erdungswiderstand, Spannungsfestigkeit und Rauheit · Originaldokument (Englisch)
Beim Vertrieb anfragen
DQ (Designqualifizierung)
Vollständiges Designqualifizierungsprotokoll · Originaldokument (Englisch)
Beim Vertrieb anfragen
Werkstoffzeugnisse
Zusammensetzung des Edelstahls 316L · Originaldokument (Englisch)
Beim Vertrieb anfragen
SED