SED-MCY-A — Sealing Tester / Leak Tester
لا تزال ترجمة الوصف الكامل لهذا الطراز قيد العمل. راجع التفاصيل بالإنجليزية أو اطلب المعلومات من فريقنا. English
SED-MCY-A is a vacuum sealing tester (leak tester) designed and manufactured per GB/T 15171-94 — test method for sealing performance of flexible packaging — widely used for QC of food, pharmaceutical, daily-chemical and
مبدأ العمل
Sample is placed inside the acrylic transparent vacuum tank with water around it(water level~10 mm above the perforated press plate). The instrument evacuates the tank via either a compressed-air-driven Venturi vacuum generator(compressor version) or an external rotary-vane vacuum pump(vacuum-pump version). When the tank pressure reaches the set negative pressure ±accuracy band, the pump cycles off; if pressure rises above the band, the pump kicks back in. Sealed packaging in the water has higher internal pressure than the surrounding tank — leaks emit a stream of bubbles visible through the acrylic. After the user-set hold time elapses(1–9999 s), the instrument automatically opens the relief solenoid valve, returning tank pressure to atmospheric.
المواصفات الفنية
المواصفات
الأبعاد
المواد
التشغيل
البيانات الفيزيائية
الخصائص الرئيسية
التطبيقات
القطاعات
العلامات التجارية للمكونات
التنظيف والصيانة
Keep air source dry and clean; regularly clean inlet filter. Drain water from air compressor regularly. Keep equipment and surroundings clean. Periodically check pressure-display accuracy; recalibrate if drift is found(calibration password 3052). Check vacuum-pump oil level — keep between upper and lower marks; replace oil when emulsified or contaminated.