Sala de exposición en Ontario, California · Consulte el soporte para su destino
Inicio / Equipos / Aislador de presión negativa / Negative-Pressure Isolator
SED-3FGL-A
Consultar
SED-3FGL-A
SED · SED-3FGL-A · Destacado

SED-3FGL-A — Negative-Pressure Isolator

La descripción completa de este modelo aún se está traduciendo. Consulte los detalles en inglés o solicite información a nuestro equipo. English

SED-3FGL-A is a negative-pressure isolator (containment isolator) configured for high-potency or hazardous-material handling. It combines a sealed main chamber with material-transfer airlock and continuous-bag (continuou

Plazo de entrega
Consultar
Capacidad de producción
Contáctenos

Funcionamiento

An EC scroll fan draws clean air through Camfil bag-in/bag-out HEPA drum filters into the main chamber, then exhausts it through PUSH-PUSH safe-change HEPA filters. By balancing supply and exhaust airflow with pneumatic butterfly valves, the controller maintains the chamber at negative pressure (monitored by Dwyer differential gauges/transmitter) so airborne dust cannot escape into the room. Operators manipulate product through gloves; product enters and leaves via the transfer airlock and continuous liner.

Proceso de funcionamiento
1An EC scroll fan draws clean air through Camfil bag-in/bag-out HEPA drum filters into the main chamber, then exhausts it through PUSH-PUSH safe-change HEPA filters.
2By balancing supply and exhaust airflow with pneumatic butterfly valves, the controller maintains the chamber at negative pressure (monitored by Dwyer differential gauges/transmitter) so airborne dust cannot escape into the room.
3Operators manipulate product through gloves; product enters and leaves via the transfer airlock and continuous liner.

Especificaciones técnicas

Especificaciones de la variante: SED-3FGL-A

Especificaciones

Modelo
SED-3FGL-A

Características principales

Negative-pressure containment isolator with main chamber and material-transfer airlock
Continuous-bag (continuous liner) discharge, 300 mm diameter
Operator gloves on POM oval flanges (8 in. oval, 8 ports)
Camfil bag-in/bag-out HEPA filtration: 2x supply drum, 1x balance, 2x PUSH-PUSH safe-change
EBM EC scroll fan G3G160-CU09-11
Pneumatic double-acting butterfly isolation valves (Chengtong, 2 in. supply x2, 2.5 in. exhaust)
Pressure-cascade monitoring with five Dwyer 0-500 Pa gauges + 1 Dwyer +/-500 Pa transmitter
Siemens S7-200 SMART PLC with analog-output module
Deli Long TPC1162HN HMI and Haide E-500 SCADA
Schneider breakers (NSC60E3020N) and Y4 relays
Mean Well NDR-240-24 24 VDC power supply
Airtac valve-island, FRL pressure regulators, solenoid valves
15 mm Enqi tempered glass doors with hinges and safety locks
Parker SG-01-1 wash and air guns for clean-in-place
Philips internal backlight
Rotadler (Germany) 8 in. oval sterile gloves
BIBO HEPA filtration plus PUSH-PUSH safe change limits operator exposure during filter swaps
EC scroll fan with PLC-coordinated dampers gives stable negative pressure across load changes
Continuous liner outlet enables closed product transfer of OEB 4-5 class compounds
All major instrumentation (Dwyer, Camfil, EBM, Siemens, Schneider, Parker, Rotadler) sourced from internationally recognized brands

Aplicaciones

Containment of high-potency APIs and toxic powders
Weighing, dispensing, sampling of OEB 4-5 class compounds
Sterile or aseptic operations requiring negative-pressure isolation
Continuous-liner unloading of hazardous powders

Sectores

Pharmaceutical containment (HPAPI, oncology, hormones)
Biopharma aseptic operations
Specialty chemicals with toxic dust exposure

Marcas de componentes

HMI touchscreen
Deli Long (TPC1162HN)
SCADA software
Haide E-500
PLC CPU module
Siemens S7-200 SMART
Analog-output module
Siemens 6ES7 231-4HF32-0XB0
Network switch
Domestic EKI-2525
Pressure regulators (FRL)
Airtac GFC30008-F1
DC power supply
Mean Well NDR-240-24
Breakers
Schneider NSC60E3020N
Solenoid valve island
Airtac 100M-11F manifold + 4V110/3V110
Relays
Schneider Y4
Supply HEPA drum filter
Camfil 3202.04.00
Balance HEPA drum filter
Camfil 3201.01.00
PUSH-PUSH safe-change filter
Camfil 3206.03.00
PUSH-PUSH housing
Camfil 160 x 500 mm
EC centrifugal scroll fan
EBM G3G160-CU09-11
Supply pneumatic butterfly valve
Chengtong 2 in.
Exhaust pneumatic butterfly valve
Chengtong 2.5 in.
Differential pressure gauges
Dwyer 2000-500Pa
Differential pressure transmitter
Dwyer (-500 to +500 Pa)
Wash gun
Parker SG-01-1
Air gun
Parker SG-01-1
Backlight panel
Philips
Tempered glass (15 mm)
Enqi
Sterile gloves (8 in. oval)
Rotadler
SED
Hable con SED Sales Team
Especialistas en equipos · contacto por WhatsApp
Consulte el horario y la disponibilidad del equipo

Documentación

Ficha técnica (PDF)
Resumen del equipo, especificaciones y dimensiones · Documento original (inglés)
Solicitar al equipo comercial
Manual del operador
Instalación, funcionamiento y diagnóstico · Documento original (inglés)
Solicitar al equipo comercial
Ficha comparativa (Estándar y Económica)
Comparación de configuraciones · Documento original (inglés)
Solicitar al equipo comercial
Protocolos IQ / OQ / PQ
Documentación de validación para conformidad GMP · Documento original (inglés)
Solicitar al equipo comercial
Protocolos FAT / SAT
Pruebas de aceptación en fábrica y en sitio · Documento original (inglés)
Solicitar al equipo comercial
Planos CAD (STEP / IGES / DWG)
Planos de ingeniería para la instalación · Documento original (inglés)
Solicitar al equipo comercial
Certificados de calibración
Nivel sonoro, resistencia de tierra, rigidez dieléctrica y rugosidad · Documento original (inglés)
Solicitar al equipo comercial
DQ (cualificación del diseño)
Protocolo completo de cualificación del diseño · Documento original (inglés)
Solicitar al equipo comercial
Certificados de materiales
Composición del acero inoxidable 316L · Documento original (inglés)
Solicitar al equipo comercial