Ausstellung in Ontario, Kalifornien · Support am Zielort anfragen
Startseite / Maschinen / Dampfsterilisator / Horizontal Pulsed-Vacuum Steam Sterilizer
SED-150ZQM-D
Auf Anfrage
SED-150ZQM-D
SED · SED-150ZQM-D · Hervorgehoben

SED-150ZQM-D — Horizontal Pulsed-Vacuum Steam Sterilizer

Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English

Horizontal pulsed-vacuum steam sterilizer (SED-150ZQM-D). Uses saturated steam to sterilize items. Round chamber, self-contained steam generator with external steam option, single door, jacketed design, main material sta

Lieferzeit
Auf Anfrage
Durchsatz
Kontaktieren Sie uns

Funktionsweise

Pulsed-vacuum steam sterilization: 3+ vacuum pulses to evacuate air from packaged, unpackaged, hollow, porous, and internally-piped instruments. Sterilization at 105-134°C with steam pressure 0.05-0.22 MPa (microcomputer-controlled). Includes timed sterilization (0-99 min) and deep drying (0-99 min). Multi-stage safety: door interlock prevents opening when chamber pressure >0.027 MPa, safety valve over-pressure relief, low-water cutoff, over-current breaker. Bowie & Dick test function for steam penetration verification. Printer logs full cycle.

Prozessbeschreibung
1Pulsed-vacuum steam sterilization: 3+ vacuum pulses to evacuate air from packaged, unpackaged, hollow, porous, and internally-piped instruments.
2Sterilization at 105-134°C with steam pressure 0.05-0.22 MPa (microcomputer-controlled).
3Includes timed sterilization (0-99 min) and deep drying (0-99 min).
4Multi-stage safety: door interlock prevents opening when chamber pressure >0.027 MPa, safety valve over-pressure relief, low-water cutoff, over-current breaker.
5Bowie & Dick test function for steam penetration verification.
6Printer logs full cycle.

Technische Daten

Technische Daten für Variante: SED-150ZQM-D

Technische Daten

Modell
SED-150ZQM-D
Spannung
AC380V 50Hz
Versorgung
9 kW / AC380V 50Hz; water supply for steam generator; steam supply if using external mode; drain connection
Konformitätsangaben
Pressure-vessel sterilizer with safety interlocks: door cannot open when chamber pressure >0.027 MPa; safety valve relieves over-pressure; low-water and over-current protection.
Leistung
9 kW

Abmessungen

Länge
1400 mm
Breite
650 mm
Höhe
1600 mm
Gewicht
340 kg
Aufstellfläche
Equipment dimensions 1400 × 650 × 1600 mm (L×W×H); transport dimensions 1560 × 820 × 1820 mm; net weight 340 kg / gross 430 kg

Werkstoffe

Kontaktwerkstoffe
Edelstahl
Gehäusewerkstoffe
Edelstahl; carbon steel option for outer shell

Wesentliche Merkmale

Capacity 150L (Φ440 × 1000 mm)
Design temperature 150°C / pressure 0.3 MPa
Rated working: 134°C / 0.22 MPa
Temperature range 105-134°C, thermal uniformity ≤±2°C
Vacuum limit -0.08 MPa with 3+ pulse pre-vacuum
Sterilization timer 0-99 min, drying timer 0-99 min
Microcomputer automatic control (no monitoring needed)
Deep drying function for sterilized items
Built-in printer logs full sterilization cycle
Bowie & Dick test function for steam penetration
Auto fault detection
Self-contained steam generator + external steam option
Jacketed single-door design
Weight 340 kg, dimensions 1400 × 650 × 1600 mm
Pulsed vacuum suits packaged, unpackaged, hollow, porous, and internally-piped instruments
Comprehensive safety: over-pressure, over-temperature, auto pressure stabilization, low-water, over-current, door interlock at 0.027 MPa
Steam pressure control 0.05-0.22 MPa with corresponding 110-134°C temperatures, selectable on control panel
Premium stainless steel sterilization chamber
Automatic discharge of air and condensate balances chamber temperature

Anwendungen

Sterilization of medical instruments
Sterilization of dressings
Sterilization of glassware
Sterilization of solution culture media
Sterilization of packaged, unpackaged, hollow, porous, and internally-piped devices

Branchen

Healthcare facilities
Hospitals
Laboratories
Pharma
SED
Sprechen Sie mit SED Sales Team
Maschinenspezialisten · Kontakt per WhatsApp
Geschäftszeiten und Verfügbarkeit beim Team erfragen

Dokumentation

Datenblatt (PDF)
Überblick, technische Daten und Abmessungen · Originaldokument (Englisch)
Beim Vertrieb anfragen
Bedienhandbuch
Einrichtung, Betrieb und Fehlersuche · Originaldokument (Englisch)
Beim Vertrieb anfragen
Vergleichsdatenblatt (Standard und Economy)
Vergleich der Konfigurationen · Originaldokument (Englisch)
Beim Vertrieb anfragen
IQ / OQ / PQ-Protokolle
Validierungsunterlagen zur GMP-Konformität · Originaldokument (Englisch)
Beim Vertrieb anfragen
FAT / SAT-Protokolle
Abnahmeprüfungen im Werk und vor Ort · Originaldokument (Englisch)
Beim Vertrieb anfragen
CAD-Zeichnungen (STEP / IGES / DWG)
Konstruktionszeichnungen für die Aufstellung · Originaldokument (Englisch)
Beim Vertrieb anfragen
Kalibrierzertifikate
Schallpegel, Erdungswiderstand, Spannungsfestigkeit und Rauheit · Originaldokument (Englisch)
Beim Vertrieb anfragen
DQ (Designqualifizierung)
Vollständiges Designqualifizierungsprotokoll · Originaldokument (Englisch)
Beim Vertrieb anfragen
Werkstoffzeugnisse
Zusammensetzung des Edelstahls 316L · Originaldokument (Englisch)
Beim Vertrieb anfragen
SED
WhatsApp · SED Vertrieb
Vertriebskontakt · WhatsApp
→