SED-8PYM-J — Culture-Medium Sterilizer
A descrição completa deste modelo ainda está sendo traduzida. Consulte os detalhes em inglês ou solicite informações à nossa equipe. English
SED-8PYM-J Culture-Medium Sterilizer (Pulse Vacuum Saturated-Steam Autoclave). Horizontal double-door pass-through pulse-vacuum autoclave with 5340×1000×1500 mm chamber (interior) for sterilizing culture media, dressings
Como funciona
Saturated steam is used as sterilizing medium. Steam condenses on cool items, releasing latent heat and humidity, coagulating microbial proteins for kill. Process: Auto-prep → Pulse vacuum (eliminates residual cold air) → Heat-up → Sterilize (134°C, regulated time) → Vacuum-dry (with jacket drying) → End. Two-temperature monitoring (chamber + liquid in container). Real-time PLC pressure/temperature regulation; F0 value calculation; trend chart printed at end of cycle.