SED-8PYM-J — Culture-Medium Sterilizer
La descripción completa de este modelo aún se está traduciendo. Consulte los detalles en inglés o solicite información a nuestro equipo. English
SED-8PYM-J Culture-Medium Sterilizer (Pulse Vacuum Saturated-Steam Autoclave). Horizontal double-door pass-through pulse-vacuum autoclave with 5340×1000×1500 mm chamber (interior) for sterilizing culture media, dressings
Funcionamiento
Saturated steam is used as sterilizing medium. Steam condenses on cool items, releasing latent heat and humidity, coagulating microbial proteins for kill. Process: Auto-prep → Pulse vacuum (eliminates residual cold air) → Heat-up → Sterilize (134°C, regulated time) → Vacuum-dry (with jacket drying) → End. Two-temperature monitoring (chamber + liquid in container). Real-time PLC pressure/temperature regulation; F0 value calculation; trend chart printed at end of cycle.