Showroom à Ontario, Californie · Assistance à confirmer pour votre destination
SED-PH-S
Sur devis
SED-PH-S pH Meter
SED · SED-PH-S · 1 unités vendues

SED-PH-S — pH Meter

La description complète de ce modèle est encore en cours de traduction. Consultez les détails en anglais ou demandez des informations à notre équipe. English

SED-PH-S benchtop pH meter with 5.7-inch high-resolution LCD. Measures pH (-2.00 to 20.00, 0.01 resolution, ±0.01 pH accuracy), mV (-2000 to 2000 mV, 0.1 mV resolution, ±0.1% or ±0.3 mV) and temperature (-5 to 110 °C, 0.

GLP-compliant data managementIP54 ingress protection
Délai de livraison
Sur devis
Cadence
Contactez-nous

Fonctionnement

A glass pH electrode (BNC/Q9 input) is immersed in the sample solution; the electrode generates a potential proportional to H+ activity. The instrument amplifies this mV signal, applies temperature compensation from a 4-pin aviation connector temperature probe (ATC) or manual setting (VTC), and converts to pH using the slope and offset derived from the most recent buffer calibration. Auto-hold senses signal stability and locks the reading; results stream to the on-board GLP-compliant 500-record memory and out via USB or RS-232.

Processus de fonctionnement
1A glass pH electrode (BNC/Q9 input) is immersed in the sample solution; the electrode generates a potential proportional to H+ activity.
2The instrument amplifies this mV signal, applies temperature compensation from a 4-pin aviation connector temperature probe (ATC) or manual setting (VTC), and converts to pH using the slope and offset derived from the most recent buffer calibration.
3Auto-hold senses signal stability and locks the reading; results stream to the on-board GLP-compliant 500-record memory and out via USB or RS-232.

Caractéristiques techniques

Caractéristiques de la variante : SED-PH-S

Caractéristiques

Modèle
SED-PH-S
Tension
AC adapter 100-240V AC input, 9V DC output
Fluides et énergie
External AC adapter accepts 100–240V AC, outputs 9V DC — no special electrical install. No water or air required.
Détails de conformité
GLP-compliant data management with date/time stamping. Buffer support for NIST, GB (Chinese national) and DIN buffer sets. IP54 housing.

Dimensions

Longueur
242 mm
Largeur
195 mm
Hauteur
68 mm
Poids
0.9 kg
Emprise au sol
Benchtop 242 × 195 × 68 mm, 900 g. Sits on any lab bench near a sample station; optional electrode stand sold separately.

Matériaux

Matériaux en contact
Glass / Ag-AgCl pH electrode (sample-contact)

Caractéristiques principales

5.7-inch high-resolution backlit LCD display
pH range -2.00 to 20.00, resolution 0.01, accuracy ±0.01 pH
mV range -2000 to 2000 mV, resolution 0.1 mV, accuracy ±0.3 mV or ±0.1%
Temperature -5 to 110 °C, resolution 0.1 °C, accuracy ±0.2 °C (electronic unit), ±0.4 °C (0–60 °C range instrument)
Automatic / manual temperature compensation (ATC / VTC)
Up to 5-point calibration with auto buffer recognition (NIST, GB, DIN)
User-customisable buffer values
Auto-hold endpoint detection (Fast / Medium / Slow), timed and continuous read modes
GLP data management with 500-result memory and date/time stamping
USB 2.0 (flash drive / PC) and RS-232 (printer) outputs
Auto electrode diagnostics with slope / offset display
IP54 splash-resistant housing
Auto shutdown 1–60 min (or off)
Compact bench unit 242 × 195 × 68 mm, 900 g
GLP-compliant data management with date/time, slope and offset — audit-ready for QC labs
5-point calibration with auto-recognition handles NIST, GB and DIN buffer sets out of the box
Both USB (PC / flash) and RS-232 (printer) outputs — fits paper and paperless workflows
IP54 splash-resistant — robust against lab spills
5.7-inch LCD is large for the price band; readable from across the bench

Applications

QC pH measurement for pharmaceutical raw materials, intermediates and finished products
pH measurement of buffers, electrolytes, infusions and dialysates
pH measurement in food, beverage, dairy and brewing labs
pH measurement in water and wastewater labs
pH measurement in cosmetic / chemical formulation labs
Educational and research pH measurement

Secteurs

Pharmaceutical manufacturing — QC labs
Food and beverage labs
Cosmetics / personal care
Chemical / petrochemical
Water and environmental
Academic / research labs
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Conformité

GLP-compliant data management
IP54 ingress protection

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)
Demander au service commercial
SED
WhatsApp · Ventes SED
Contact commercial · WhatsApp
→