The SED-HWSX-B Constant-Temperature Water Bath is a precision lab thermostat designed for the JS-II Bloom-strength tester and other laboratory instruments. A microprocessor-controlled heater plus circulation pump holds b

Choisir une variante (1)
Délai de livraison
Sur devis
Cadence
Contactez-nous

Fonctionnement

A microprocessor reads temperature from an immersed sensor and modulates the heater on/off relative to a preset value (default 10.0 °C). A circulation pump moves bath fluid through an external chiller (ZL-III refrigeration unit) and back via inlet/outlet ports on the rear panel. Above 8 °C distilled water is recommended; below 8 °C a distilled water plus laboratory-grade ethylene glycol mixture is required (automotive antifreeze is prohibited). The system holds bath temperature within ±0.1 °C (5–30 °C range) and ±0.3 °C (30–70 °C range).

Processus de fonctionnement
1A microprocessor reads temperature from an immersed sensor and modulates the heater on/off relative to a preset value (default 10.0 °C).
2A circulation pump moves bath fluid through an external chiller (ZL-III refrigeration unit) and back via inlet/outlet ports on the rear panel.
3Above 8 °C distilled water is recommended; below 8 °C a distilled water plus laboratory-grade ethylene glycol mixture is required (automotive antifreeze is prohibited).
4The system holds bath temperature within ±0.1 °C (5–30 °C range) and ±0.3 °C (30–70 °C range).

Caractéristiques techniques

Caractéristiques de la variante : SED-HWSX-B 5–70 °C

Caractéristiques

Modèle
SED-HWSX-B
Tension
220V/50Hz (+/- 10 percent), single phase
Fluides et énergie
220 V ±10% AC, 50 Hz, max 1000 W, single phase, reliably grounded. 5 A fuse on input (spare included). AC voltage stabilizer recommended where mains supply is unstable. Distilled water (above 8 °C) or distilled water + lab-grade ethylene glycol (below 8 °C) for circulation fluid.
Détails de conformité
Designed for use at 18–28 °C ambient, 20–80% RH; storage −10 to +50 °C. Requires national-standard single-phase three-prong grounded outlet (220 V ±10%). 1-year warranty from purchase invoice date. Transport: do not tilt, invert, or subject to impact.
Puissance
1 kW

Dimensions

Longueur
580 mm
Largeur
440 mm
Hauteur
215 mm
Poids
23 kg
Emprise au sol
580×440×215 mm. Place on a solid, level bench with anti-vibration pad; minimum 0.5 m clearance on all sides for ventilation and heat dissipation. Level using built-in foot adjusters with included spirit level (longitudinal and transverse).

Caractéristiques principales

Preset temperature range 5–70 °C (±0.1 °C accuracy 5–30 °C, ±0.3 °C accuracy 30–70 °C)
Microprocessor (single-chip computer) precision temperature control
Built-in 1000 W heater + circulation pump + insulated bath tank
Cools 30 °C ambient water to 10 °C in under 45 minutes (when paired with ZL-III chiller)
Continuous operation greater than 24 hours
3-digit LCD displays both preset and actual bath temperature
Heat / control / abnormal indicator lights with audible buzzer feedback
Inlet and outlet ports on rear panel for closed-loop fluid circulation
Rear-panel AC outlet powers paired refrigeration unit
5 A input fuse with spare fuse included
Precision temperature control to ±0.1 °C for sensitive lab instruments
Designed as drop-in companion to JS-II Bloom-strength tester (gelatin/cosmetic QC)
Combined with ZL-III chiller delivers low-temperature water bath workflow in under 45 minutes
Compact bench-top footprint (580×440×215 mm) for crowded lab benches
Standard accessories include 16 freeze cups, mercury thermometer, spirit level, suction bulb
1-year warranty with free repair (excluding water/fire/earthquake damage)

Applications

Bloom-strength gelatin testing (paired with JS-II tester)
Constant-temperature sample conditioning for lab instruments
Low-temperature water bath for material testing
QC sample equilibration in pharma, food and chemical labs

Secteurs

Pharmaceutical QC laboratories
Food testing (gelatin, edible film QC)
Cosmetic and gelatin manufacturing
Chemical analytical labs
Material testing labs
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Nettoyage et entretien

Clean enclosure and bath tank monthly to control dust. Replace circulation fluid every 3 months at minimum. Do not use steel brushes (will damage tank). Drain via included hose with suction-bulb priming; refill and bleed air per manual before restart. Never run circulation pump dry. Bleed air after every fluid change.

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)
Demander au service commercial
SED
WhatsApp · Ventes SED
Contact commercial · WhatsApp
→