Showroom à Ontario, Californie · Assistance à confirmer pour votre destination
SED-2XDG-B
Sur devis
SED-2XDG-B Vacuum Freeze Dryer
SED · SED-2XDG-B · 1 unités vendues

SED-2XDG-B — Vacuum Freeze Dryer

La description complète de ce modèle est encore en cours de traduction. Consultez les détails en anglais ou demandez des informations à notre équipe. English

SED-2XDG-B is a large-scale pharmaceutical vacuum freeze dryer with 8+1 SUS304 shelves (500 × 500 mm each, 2 m² total), built-in hydraulic stoppering system, and SIEMENS PLC+DELL PC control. Shelf temperature range: -55

FDACE
Délai de livraison
Sur devis
Cadence
10–30 kg/batch

Fonctionnement

Product loaded on SUS304 shelves is frozen by silicon-oil (Dow Corning USA) circulation cooled by a 2-stage cascade refrigeration system (DANFOSS MTZ-050 + NTZ-068, France). The condenser (min ≤-75 °C) captures sublimated water vapor from the rectangular SUS304 drying chamber under vacuum (ultimate ≤1 Pa) maintained by a Japan ULVAC rotary-vane vacuum pump (18 m³/h). Shelf heating uses PID-controlled electric heating (3 kW) for precise temperature ramping. A Germany FV butterfly valve isolates the drying chamber from the condenser. Built-in REXROTH (Germany) hydraulic system enables in-situ stoppering of vials under vacuum. SIEMENS PLC + DELL PC with WINCC software supports automatic and full manual emergency operation, recipe management, and remote diagnostics.

Processus de fonctionnement
1Product loaded on SUS304 shelves is frozen by silicon-oil (Dow Corning USA) circulation cooled by a 2-stage cascade refrigeration system (DANFOSS MTZ-050 + NTZ-068, France).
2The condenser (min ≤-75 °C) captures sublimated water vapor from the rectangular SUS304 drying chamber under vacuum (ultimate ≤1 Pa) maintained by a Japan ULVAC rotary-vane vacuum pump (18 m³/h).
3Shelf heating uses PID-controlled electric heating (3 kW) for precise temperature ramping.
4A Germany FV butterfly valve isolates the drying chamber from the condenser.
5Built-in REXROTH (Germany) hydraulic system enables in-situ stoppering of vials under vacuum.
6SIEMENS PLC + DELL PC with WINCC software supports automatic and full manual emergency operation, recipe management, and remote diagnostics.

Caractéristiques techniques

Caractéristiques de la variante : SED-2XDG-B

Caractéristiques

Modèle
SED-2XDG-B
Cadence
10–30 kg/batch
Tension
380V±10%,50Hz,3Phase,5Wire
Fluides et énergie
Total power 8.65 kW, 380 V±10%, 50 Hz, 3-phase, 5-wire. Compressed air: ≥20 L/min at 0.5–0.8 MPa. Cooling water: ≥5 m³/h at ≤30 °C, 0.2–0.3 MPa.
Détails de conformité
FDA and EU/CE compliant. Supports electronic signature and multiple password levels. IQ/OQ/FAT/SAT documentation packages available. Vacuum vessel design per standard.
Puissance
8.65 kW
Pression d’air
0.5~0.8Mpa

Dimensions

Longueur
1500 mm
Largeur
1500 mm
Hauteur
2300 mm
Poids
1500 kg
Emprise au sol
1500 × 1500 × 2300 mm (L × W × H), approx. 1500 kg

Matériaux

Matériaux en contact
SUS304
Matériaux du bâti
SUS304

Caractéristiques principales

Total power: 8.65 kW (380 V±10%, 50 Hz, 3-phase, 5-wire)
Overall dimensions: 1500 × 1500 × 2300 mm
Machine weight: approx. 1500 kg
Shelf count: 8 + 1 (compensation shelf)
Shelf dimensions: 500 × 500 mm / δ20 mm
Total shelf area: 2 m²
Shelf spacing: 80 mm
Shelf material: SUS304, surface Ra ≤0.4 μm
Shelf temperature range: -55 °C to +80 °C
Shelf temperature uniformity: ≤±1 °C
Shelf flatness: ≤0.5 mm/m
Max water capture: ≥40 kg/batch
Condenser minimum temperature: ≤-75 °C
Condenser: horizontal cylinder, SUS304, external coil, Ra ≤0.8 μm
Ultimate vacuum: ≤1 Pa
Vacuum pump: Japan ULVAC rotary-vane, 18 m³/h, 0.75 kW
Vacuum degree: auto-controlled, 0–200 Pa adjustable
Refrigeration: DANFOSS MTZ-050 (3 kW, R404a) + NTZ-068 (1.5 kW, R23), cascade
Circulation pump: WILO TOP32 (Germany), 0.4 kW
Heat transfer medium: Dow Corning silicon oil (USA)
Shelf heating rate (no load): ≥1 °C/min
Shelf cooling rate (no load): 20 °C → -40 °C ≤60 min
Condenser cooling rate (no load): 20 °C → -40 °C ≤30 min
Hydraulic stoppering: REXROTH (Germany), 0–1.0 bar adjustable
Control: SIEMENS PLC + DELL PC, WINCC software
FDA/EU compliant, electronic signature, multi-password, multilingual interface
Remote diagnostics and control
Compressed air: ≥20 L/min at 0.5–0.8 MPa
Cooling water: ≥5 m³/h at T ≤30 °C, 0.2–0.3 MPa
FDA and CE certified — ready for regulated pharmaceutical markets
SIEMENS PLC + DELL PC with WINCC: recipe management, data logging, electronic signature, remote diagnostics
In-situ hydraulic stoppering (REXROTH Germany) for aseptic vial sealing under vacuum
Cascade DANFOSS (France) refrigeration to ≤-75 °C condenser and ≤1 Pa ultimate vacuum
±1 °C shelf uniformity and ≤0.5 mm/m flatness for consistent lyophilization results
Multi-language one-button interface switch for global operators
PID electric heating — energy-efficient, stable temperature control
WILO (Germany) leak-free, low-noise circulation pump

Applications

Large-scale pharmaceutical freeze-drying (vials, ampoules, bulk)
Biological products and vaccines lyophilization
API and formulation freeze-drying
GMP-compliant production freeze-drying

Secteurs

Pharmaceutique
Biotechnology
Vaccines

Marques des composants

1st-stage compressor
DANFOSS MTZ-050
2nd-stage compressor
DANFOSS NTZ-068
Oil separator
EMERSON
Pressure protection relay
Danfoss
Refrigerant sight glass
Danfoss
Dry filter
Danfoss
Expansion valve
Danfoss
Solenoid valve
Danfoss
1st water condenser
WEIYI
2nd evaporative condenser
WEIYI
Safety valve
CASTEL
Vacuum pump
ULVAC
Vacuum sensor
ULVAC SW1-N1
Vacuum pipeline valve
FV DN25
Chamber isolation valve
FV DN150
Vacuum check valve
SIEGV GYC-JQ40
Condenser drain valve
SIEGV GDC-10
Vacuum adjust valve
SIEGV GDC-8
Circulation pump
WILO TOP32
Hydraulic units
REXROTH
PLC
SIEMENS
PC
DELL
Circuit breaker
SIEMENS
AC contactor
SIEMENS
Pipe insulation
ARMAFLEX
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Conformité

FDA
CE

Nettoyage et entretien

SUS304 surfaces polished to Ra ≤0.4 μm (shelves) and Ra ≤0.8 μm (condenser). Chamber, shelves, and hoses tested through multiple procedures to ensure no leakage. Condenser interior cleaned by special method to ensure no dirt. Hot refrigerant defrost for condenser.

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)
Demander au service commercial
SED
WhatsApp · Ventes SED
Contact commercial · WhatsApp
→