Showroom à Ontario, Californie · Assistance à confirmer pour votre destination
SED-200SCD-Y
Sur devis
SED-200SCD-Y
SED · SED-200SCD-Y · 2 unités vendues

SED-200SCD-Y — Moisture Analyzer

La description complète de ce modèle est encore en cours de traduction. Consultez les détails en anglais ou demandez des informations à notre équipe. English

The SED-200SCD-Y is a halogen-lamp thermogravimetric moisture analyzer for rapid and precise determination of moisture content in pharmaceutical, food, grain, tobacco, paper, textile, mineral, and chemical raw material s

Accuracy Level 2 (per Chinese balance standard)
Délai de livraison
Sur devis
Cadence
Contactez-nous

Fonctionnement

Halogen-lamp thermogravimetric drying: place sample on tared scale pan in the heated chamber; ring halogen lamp heats sample rapidly and uniformly while built-in precision balance continuously measures weight; firmware calculates and displays % moisture loss in real time. Drying continues until either timed completion or auto-mode (weight loss per unit time falls below threshold), then the final moisture value is locked and displayed. Touchscreen menus support weight calibration (single-point with 100 g weight or 4-point linearity), clock setting, AUTO DRY (unknown samples), TIME DRY (timed), RECORD (unit conversions), and DATABASE (up to 18 saved records).

Processus de fonctionnement
1Halogen-lamp thermogravimetric drying: place sample on tared scale pan in the heated chamber; ring halogen lamp heats sample rapidly and uniformly while built-in precision balance continuously measures weight; firmware calculates and displays % moisture loss in real time.
2Drying continues until either timed completion or auto-mode (weight loss per unit time falls below threshold), then the final moisture value is locked and displayed.
3Touchscreen menus support weight calibration (single-point with 100 g weight or 4-point linearity), clock setting, AUTO DRY (unknown samples), TIME DRY (timed), RECORD (unit conversions), and DATABASE (up to 18 saved records).

Caractéristiques techniques

Caractéristiques de la variante : SED-200SCD-Y

Caractéristiques

Modèle
SED-200SCD-Y
Tension
AC 220V 50Hz (or 120V 60Hz, 1-phase) - max 480 W
Fluides et énergie
AC 220V 50Hz (or 120V 60Hz), single-phase. Max 480 W during heating. Stable mains voltage required (no exceeding limits). Operating environment: 15-30 C, fluctuation <5 C/h, 50-85% RH (non-condensing). Avoid corrosive gas (H2S, NOx fluorides), strong magnetic/electric fields, direct sunlight or strong wind. Preheating 30 min after power-on recommended.
Détails de conformité
Frequent power-on recommended to keep instrument in stable thermal state and reduce preheat time. Periodic calibration check required. Operates per halogen thermogravimetric moisture-determination standards.
Puissance
0.48 kW

Dimensions

Longueur
360 mm
Largeur
215 mm
Hauteur
185 mm
Poids
4.5 kg
Emprise au sol
Desktop unit, 360 x 215 mm footprint, 170-185 mm height; 4.5 kg net weight. Place on solid stable working platform with at least 1 m clearance above for heat dissipation. Ground connection required.

Matériaux

Matériaux en contact
metal sample plate (90 mm diameter)
Matériaux du bâti
metal shell with halogen-resistant chamber

Composants

Output Interface
RS232C

Alimentation électrique

Max Power
480 W

Fonctionnement

Drying Technology
/Ring Halogen Lamp
Max Capacity
110 g
Moisture Accuracy
±0.5 (sample ≥3g) %
Moisture Range High
100 %
Moisture Range Low
0 %
Moisture Readability
0.01 %
Balance Readability
0.001 g
Repeatability
0.1 %
Temperature Control Accuracy
±1 °C
Temperature Range Max
200 °C
Time Control Max
99 min

Caractéristiques physiques

Pan Size
90 mm

Caractéristiques principales

Maximum sample weight 110 g
Balance readability 0.001 g (1 mg); calibration scale 10 mg
Moisture readability 0.01% (or 0.02-0.1% for samples >=3 g)
Moisture measurement accuracy +/-0.5% (sample >=3 g)
Repeatability 0.1%
Moisture content range 0-100%
Temperature range room temp to 200 C; accuracy +/-1 C
Heating time 0-99 min (1-min increments)
Ring halogen lamp heating element (rapid, uniform)
Color TFT touchscreen HMI with 9 display modes (moisture, dry weight, ATRO dry, ATRO wet, etc.)
Stop modes: timed (manual) or auto
RS232C output for PC / printer (9600 baud)
Single-point and 4-point linearity calibration
Up to 18 records saved in onboard database
End-of-heating buzzer
Sample plate diameter 90 mm
AC 220V 50Hz (also 120V 60Hz available); max 480 W
Fast: typical samples measured in just a few minutes
High precision: 1 mg balance + +/-0.5% moisture accuracy meets pharma/QC needs
Touchscreen with 9 display modes - flexible for different reporting requirements
RS232C interface for PC/printer integration in QC labs
Compact desktop unit (4.5 kg) - portable across labs
Unique high-precision insulated sensor for reliable temperature data

Applications

Free water / moisture content determination in raw materials
QC sampling in pharma manufacturing (granulation moisture, etc.)
Food, grain, tobacco, paper, textile moisture testing
Mineral and chemical raw material moisture analysis
Laboratory R&D moisture studies

Secteurs

Pharmaceutique
Food and grain processing
Tobacco
Paper
Textile
Mineral / mining
Chemical raw materials
Laboratory and R&D

Marques des composants

Imported halogen heating lamp
High-precision insulated thermal sensor
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Conformité

Accuracy Level 2 (per Chinese balance standard)

Nettoyage et entretien

Keep clean and dust-free. Clean scale pan and shell with detergent and soft cloth; alcohol and gasoline are forbidden. Periodic check and recalibration recommended. After each measurement, allow chamber to cool to room temperature (open cover) before next sample.

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)
Demander au service commercial
SED
WhatsApp · Ventes SED
Contact commercial · WhatsApp
→