SED-1000MM-U pulsating vacuum pressure steam sterilizer (PVS series). Cabinet style. Effective chamber volume >=1000 L; inner chamber dimensions 610 x 910 x 1820 mm. Double-leaf motorized door with single-door single-sea

GB8599-2008WS 310 (hospital sterilization)ISO9001ISO13485ISO14001ISO45001
Délai de livraison
Sur devis
Cadence
1.2–1.5 L (chamber volume)

Fonctionnement

Multiple negative-pressure, cross-pressure and positive-pressure pulses (>=3, adjustable 0-99) evacuate cold air from the load; saturated steam is then admitted to reach 121 degrees C or 134 degrees C and held for the programmed time (20 min at 121 / 5 min at 134, settable 0-180 min). After sterilization, vacuum drying combined with pulsating air drying removes residual moisture (drying time 10 min, settable 0-180 min). Siemens PLC controls the pulse sequences, temperature, pressure and drying via pneumatic angle-seat valves. The built-in 60 kW steam generator (design pressure 0.7 MPa) provides the steam source.

Processus de fonctionnement
1Multiple negative-pressure, cross-pressure and positive-pressure pulses (>=3, adjustable 0-99) evacuate cold air from the load; saturated steam is then admitted to reach 121 degrees C or 134 degrees C and held for the programmed time (20 min at 121 / 5 min at 134, settable 0-180 min).
2After sterilization, vacuum drying combined with pulsating air drying removes residual moisture (drying time 10 min, settable 0-180 min).
3Siemens PLC controls the pulse sequences, temperature, pressure and drying via pneumatic angle-seat valves.
4The built-in 60 kW steam generator (design pressure 0.7 MPa) provides the steam source.

Caractéristiques techniques

Caractéristiques de la variante : SED-1000MM-U

Caractéristiques

Modèle
SED-1000MM-U
Cadence
1.2–1.5 L (chamber volume)
Tension
~380V, 50Hz
Fluides et énergie
3-phase 380V/50Hz (total 63.5 kW); built-in steam generator or external steam supply; cooling water; compressed air for door seal and valves
Détails de conformité
Complies with GB8599-2008 'Technical Requirements for Large Steam Sterilizers' and hospital WS 310 series; manufacturer holds ISO9001, ISO13485, ISO14001, ISO45001. Air leakage <=0.13 kPa/min; dynamic pressure change <=1000 kPa/min per any 3-second interval.
Puissance
63.5 kW
Pression d’air
Interior design pressure -0.1 to 0.3 MPa; jacket design pressure 0.3 MPa; steam generator design pressure 0.7 MPa

Dimensions

Longueur
1820 mm
Largeur
1440 mm
Hauteur
2370 mm
Poids
1580 kg
Emprise au sol
Width 1440 mm (DB golden); inner chamber width x length 610 x 1820 mm

Matériaux

Matériaux en contact
SUS30408 stainless steel (inner chamber and door); sanitary electrochemically polished SS304 pipework; silicone door seal
Matériaux du bâti
Inner shell SUS30408; jacket Q245R; outer panel 304 frosted anti-fingerprint SS; aluminum-silicate fibre insulation >=60 mm; movable casters

Caractéristiques principales

Effective chamber volume >=1000 L; inner chamber 610 x 910 x 1820 mm
Sterilization temperature 121 and 134 degrees C, settable 115-138 degrees C
Sterilization time: 20 min @ 121 / 5 min @ 134; settable 0-180 min
Standard cycle <=55 min
Pulse frequency >=3, adjustable 0-99; negative + cross + positive pressure pulses
Vacuum drying + pulsating air drying; drying time settable 0-180 min
Double-leaf motorized door, single-door single-seal, front-rear sealing interlock
Inner shell SUS30408; jacket Q245R; European-style ring-reinforced (GETINGE-type) design
Plasma-welded inner shell; robot-welded jacket (documentation provided)
Built-in 60 kW electric steam generator (design pressure 0.7 MPa)
German SPECK / Shanghai Yulong direct-coupled water-ring vacuum pump; plate heat exchanger pre-cool; ~30% cooling water recirculation
Pneumatic angle-seat valves rated >4 million cycles; sanitary SS tri-clamp pipework
Total installed power 63.5 kW (3.5 kW control + 60 kW steam generator)
Siemens PLC; dual 10 inch colour HMI front and rear; built-in thermal printer (5-year archive); optional permanent electronic record
Network protocol support for remote operation; interface to information management and traceability systems
Programs: textiles, instruments, thermal load, fast cycle, orthopaedic, prions, lumen, heavy/small loads, open liquids, gravity, BD test, leak test, vacuum drying
Chlorine- and asbestos-free aluminum-silicate fibre insulation >=60 mm; outer surface temp <= ambient + 25 degrees C
Outer panel 304 frosted anti-fingerprint stainless steel
Long-life silicone door seal: >=1500 cycles / 2-year design life
Air leakage pressure rise <=0.13 kPa/min
Dynamic pressure change <=1000 kPa/min per any 3-second interval during sterilization
Phase-sequence protection; alarm system; movable casters
Designed life 10 years / 20000 cycles (optional 15 years / 30000 cycles)
Manufacturer: ISO9001, ISO13485, ISO14001, ISO45001
Net weight 1580 kg
Double-door sterile/dirty barrier for CSSD and clean-room applications
European ring-reinforced jacket for superior fatigue life under pulsating pressure loads
Robot welding eliminates human-induced weld defects (proof documentation provided)
Connectable to hospital/pharma quality traceability and information management systems
Energy-saving non-asbestos insulation; surface temperature near ambient
Long-life silicone door seal (>=1500 cycles)

Applications

Sterilization of textiles, dressings, instruments, glassware
Sterilization of hollow/lumen instruments and injection equipment
Sterilization of solids, liquids and culture media
Pharmaceutical injectable manufacturing sterilization

Secteurs

Pharmaceutical (injectable equipment)
Hospital / medical CSSD
Food
Scientific research / safety laboratories

Marques des composants

Siemens (Germany) PLC
German SPECK / Shanghai Yulong water-ring vacuum pump
WIKA (Germany) pressure gauges and transmitters
International-brand 10 inch colour HMI
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Conformité

GB8599-2008
WS 310 (hospital sterilization)
ISO9001
ISO13485
ISO14001
ISO45001

Nettoyage et entretien

Inner-chamber floor inclined toward drain for easy self-draining and cleaning. Condensate drained automatically with drainage temperature regulated to 55 degrees C.

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)