Sala de exposición en Ontario, California · Consulte el soporte para su destino
Inicio / Equipos / Esterilizador a vapor / Steam Sterilizer (Tabletop Pre-Vacuum, Class B)
SED-23ZQM-U
Consultar
SED-23ZQM-U
SED · SED-23ZQM-U · Destacado

SED-23ZQM-U — Esterilizador a vapor

La descripción completa de este modelo aún se está traduciendo. Consulte los detalles en inglés o solicite información a nuestro equipo. English

SED-23ZQM-U is a 23 L tabletop pre-vacuum pressure-steam sterilizer (Class II medical device). It uses electric heating and multiple pulsating vacuums for fast heat-up and thorough sterilization, with powerful vacuum dry

Class II medical deviceAir leak rate ≤0.13 kPa/min (EN 13060-aligned)
Plazo de entrega
Consultar
Capacidad de producción
Contáctenos

Funcionamiento

Distilled water (~0.5–1 L) feeds from the 2 L tank into the electric-heated chamber. The double-head oil-free vacuum pump performs multiple pulsating pre-vacuum cycles (Class B-style) to evacuate air from the Φ246 × 450 mm chamber and load. Saturated steam at 121 °C or 134 °C is then injected; pressure rises to 0.21 MPa with humidity ≥80%. The microcomputer holds for the user-set time, monitored by built-in pressure and temperature sensors. Powerful post-vacuum drying removes moisture (residual <0.2% on instruments, <1% on dressings). Waste steam is condensed in the built-in collector and the cycle ends without external steam release. Micro-printer logs minute-by-minute data; high/low alarms and overpressure protections trigger as needed.

Proceso de funcionamiento
1Distilled water (~0.5–1 L) feeds from the 2 L tank into the electric-heated chamber.
2The double-head oil-free vacuum pump performs multiple pulsating pre-vacuum cycles (Class B-style) to evacuate air from the Φ246 × 450 mm chamber and load.
3Saturated steam at 121 °C or 134 °C is then injected; pressure rises to 0.21 MPa with humidity ≥80%.
4The microcomputer holds for the user-set time, monitored by built-in pressure and temperature sensors.
5Powerful post-vacuum drying removes moisture (residual <0.2% on instruments, <1% on dressings).
6Waste steam is condensed in the built-in collector and the cycle ends without external steam release.
7Micro-printer logs minute-by-minute data; high/low alarms and overpressure protections trigger as needed.

Especificaciones técnicas

Especificaciones de la variante: SED-23ZQM-U

Especificaciones

Modelo
SED-23ZQM-U
Tensión
AC 220V 50Hz
Potencia
1.7 kW

Dimensiones

Longitud
715 mm
Anchura
513 mm
Altura
425 mm
Peso
49 kg

Materiales

Materiales en contacto
304 anti-corrosion stainless steel sterilization chamber; silicone door seal ring
Materiales del cuerpo
ABS outer casing; 304 stainless steel chamber

Características principales

23 L pre-vacuum pressure-steam sterilizer
Multiple pulsating vacuum cycles (faster heat-up, thorough sterilization)
Powerful vacuum drying — instrument residual humidity <0.2%, dressing <1%
6 sterilization modes: bare/wrapped instrument, fabric/hollow, rubber, B&D test, vacuum test
Microcomputer intelligent control with automatic door lock
LCD temperature display
Chamber Φ246 × 450 mm in 304 anti-corrosion stainless steel
ABS outer casing
Double-head oil-free vacuum pump (leather cup)
Built-in pressure sensor
Built-in temperature sensor
Micro-printer (minute-by-minute permanent record)
Built-in condensate collector — internal steam recycling, no external venting
Automatic water-quality monitoring
Silicone seal ring (2-year / 1,200-cycle service life)
Safety valve auto-opens at 250 kPa
Overpressure auto-vent at 220 kPa
Overpressure program-stop and auto-release at 230 kPa
Air leak rate ≤0.13 kPa/min at each test stage
Dynamic chamber pressure change ≤1000 kPa/min in any 2 s
High/low temperature alarm system
Energy consumption ~1.5 kWh per cycle
Power 1.7 kVA, AC 220V 50Hz
Dimensions 715 × 513 × 425 mm, weight 49 kg
Class II medical device

Aplicaciones

Surgical, dental and ophthalmic instrument sterilization
Wrapped/encapsulated instrument sterilization
Fabric and hollow load sterilization
Rubber product sterilization
B&D and vacuum-leak validation testing

Sectores

Hospitals and medical clinics
Dental clinics
Pharmaceutical R&D
Laboratorio
Beauty / aesthetic clinics
SED
Hable con SED Sales Team
Especialistas en equipos · contacto por WhatsApp
Consulte el horario y la disponibilidad del equipo

Conformidad

Class II medical device
Air leak rate ≤0.13 kPa/min (EN 13060-aligned)

Documentación

Ficha técnica (PDF)
Resumen del equipo, especificaciones y dimensiones · Documento original (inglés)
Solicitar al equipo comercial
Manual del operador
Instalación, funcionamiento y diagnóstico · Documento original (inglés)
Solicitar al equipo comercial
Ficha comparativa (Estándar y Económica)
Comparación de configuraciones · Documento original (inglés)
Solicitar al equipo comercial
Protocolos IQ / OQ / PQ
Documentación de validación para conformidad GMP · Documento original (inglés)
Solicitar al equipo comercial
Protocolos FAT / SAT
Pruebas de aceptación en fábrica y en sitio · Documento original (inglés)
Solicitar al equipo comercial
Planos CAD (STEP / IGES / DWG)
Planos de ingeniería para la instalación · Documento original (inglés)
Solicitar al equipo comercial
Certificados de calibración
Nivel sonoro, resistencia de tierra, rigidez dieléctrica y rugosidad · Documento original (inglés)
Solicitar al equipo comercial
DQ (cualificación del diseño)
Protocolo completo de cualificación del diseño · Documento original (inglés)
Solicitar al equipo comercial
Certificados de materiales
Composición del acero inoxidable 316L · Documento original (inglés)
Solicitar al equipo comercial
SED