SED-2BJY-B Disintegration Tester
SED · SED-2BJY-B · 5 verkaufte Einheiten

SED-2BJY-B — Disintegration Tester

Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English

SED-2BJY-B is a benchtop dual-basket disintegration tester for tablets, hard/soft capsules, and pills per pharmacopoeia method. Two independent baskets each hold 6 glass tubes; baskets are raised/lowered at 30–31 cycles/

Variante wählen (1)
Lieferzeit
Auf Anfrage
Durchsatz
30–31 times/min

Funktionsweise

Six pharmacopoeia-compliant glass tubes are mounted in each of two independent baskets and immersed in a 1000 mL beaker placed in a constant-temperature water bath (heater + circulation pump). The MPU controller drives the basket lift mechanism to move baskets up and down at 30–31 cycles/min with 55±1 mm stroke. Each interruption of the tablet's passage through the mesh is timed automatically. High-precision electronic temperature sensors monitor the bath and display current vs. preset temperature; an over-temperature alarm and auto-protection circuit cut the heater if temperature exceeds the preset limit.

Prozessbeschreibung
1Six pharmacopoeia-compliant glass tubes are mounted in each of two independent baskets and immersed in a 1000 mL beaker placed in a constant-temperature water bath (heater + circulation pump).
2The MPU controller drives the basket lift mechanism to move baskets up and down at 30–31 cycles/min with 55±1 mm stroke.
3Each interruption of the tablet's passage through the mesh is timed automatically.
4High-precision electronic temperature sensors monitor the bath and display current vs.
5preset temperature; an over-temperature alarm and auto-protection circuit cut the heater if temperature exceeds the preset limit.

Technische Daten

Technische Daten für Variante: SED-2BJY-B dual-basket disintegration tester (16 kg

Technische Daten

Modell
SED-2BJY-B
Durchsatz
30–31 times/min
Spannung
AC220V/50Hz
Versorgung
AC 220V ±10% / 50Hz, max 600 W total (heater 500 W). Distilled water as bath cycle liquid. 3A fuse (backup fuse supplied).
Konformitätsangaben
Built to pharmacopoeia disintegration-test method (37.0 °C default, 30–31 cycles/min, 55±1 mm stroke). Over-temperature alarm and auto-protection included. 12-month warranty from installation; lifetime maintenance service.
Leistung
0.6 kW

Abmessungen

Länge
400 mm
Breite
320 mm
Höhe
440 mm
Gewicht
16 kg
Aufstellfläche
Benchtop instrument, 400 mm L × 320 mm W × 440 mm H, 16 kg net. Requires ~0.5 m clearance around the unit for heat dissipation. Place on stable workbench; use damping pad between bath box and tester to reduce vibration.

Werkstoffe

Kontaktwerkstoffe
Borosilicate glass tubes; stainless-steel basket frames and discs; PTFE/plastic seals
Gehäusewerkstoffe
Painted steel housing with stainless-steel water-bath tank

Elektrik

Heater Power
500 W

Betrieb

Basket Lifting Frequency High
31 cycles/min
Basket Lifting Frequency Low
30 cycles/min
Basket Lifting Stroke
55±1 mm
Preset Temperature Max
50 °C
Preset Temperature Min
20 °C
Automatic Temp Accuracy
±0.5 °C
Temperature Control Accuracy
±0.3 °C
Temperature Resolution
0.1 °C
Timing Accuracy
±0.5 min
Timing Range Max
900 min
Timing Range Min
1 min
Working Noise
<60 dB

Physikalische Angaben

Standard Mesh Aperture
2 mm
Min gap mesh-to-beaker-bottom
25±2 mm
Number of Baskets
2 sets

warranty

Warranty
12 months

Wesentliche Merkmale

Dual independent baskets — 2 tests in parallel
Each basket holds 6 pharmacopoeia glass tubes
Basket lift frequency: 30–31 cycles/min
Lift stroke: 55±1 mm
1000 mL beaker with constant-temperature water bath
Temperature range: 20–50 °C, default 37.0 °C (pharmacopoeia standard)
Temperature control accuracy: ±0.3 °C, resolution 0.1 °C
Timing range: 1–900 min, accuracy ±0.5 min
MPU (microprocessor) central control
Over-temperature alarm and auto-protection
Working noise: <60 dB
Standard mesh aperture 2 mm; optional 0.425 mm, 0.71 mm, 1 mm
Pharmacopoeia-compliant default settings (37.0 °C, 15 min, 30–31 cycles/min) for tablet/capsule QC
Two independent baskets allow parallel testing or QA + sample comparison
Compact benchtop footprint (400 × 320 × 440 mm, 16 kg)
Calibration-free high-precision digital temperature sensor
One-year warranty, lifetime maintenance support

Anwendungen

Disintegration time testing of tablets per pharmacopoeia
Disintegration time testing of hard/soft capsules
QC release testing in pharmaceutical labs
Stability and formulation development testing

Branchen

Pharmaceutical QC and R&D
Nutraceutical / health products QC
Hospital pharmacy QC
University / research labs
SED
Sprechen Sie mit SED Sales Team
Maschinenspezialisten · Kontakt per WhatsApp
Geschäftszeiten und Verfügbarkeit beim Team erfragen

Reinigung und Wartung

Clean basket (nacelle) units after each test to avoid build-up. Periodically clean the bath box and all parts; never use steel brushes — they damage surfaces. Empty bath box by siphoning liquid out before cleaning. Wipe down all exterior surfaces.

Dokumentation

Datenblatt (PDF)
Überblick, technische Daten und Abmessungen · Originaldokument (Englisch)
Beim Vertrieb anfragen
Bedienhandbuch
Einrichtung, Betrieb und Fehlersuche · Originaldokument (Englisch)
Beim Vertrieb anfragen
Vergleichsdatenblatt (Standard und Economy)
Vergleich der Konfigurationen · Originaldokument (Englisch)
Beim Vertrieb anfragen
IQ / OQ / PQ-Protokolle
Validierungsunterlagen zur GMP-Konformität · Originaldokument (Englisch)
Beim Vertrieb anfragen
FAT / SAT-Protokolle
Abnahmeprüfungen im Werk und vor Ort · Originaldokument (Englisch)
Beim Vertrieb anfragen
CAD-Zeichnungen (STEP / IGES / DWG)
Konstruktionszeichnungen für die Aufstellung · Originaldokument (Englisch)
Beim Vertrieb anfragen
Kalibrierzertifikate
Schallpegel, Erdungswiderstand, Spannungsfestigkeit und Rauheit · Originaldokument (Englisch)
Beim Vertrieb anfragen
DQ (Designqualifizierung)
Vollständiges Designqualifizierungsprotokoll · Originaldokument (Englisch)
Beim Vertrieb anfragen
Werkstoffzeugnisse
Zusammensetzung des Edelstahls 316L · Originaldokument (Englisch)
Beim Vertrieb anfragen