SED-15CCS-Q — Laboratory Ultrapure Water System
Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English
Laboratory ultrapure water system (SED-15CCS-Q), 10-15 L/h, produces water meeting Grade II laboratory standard. Inlet: municipal tap water. Dual outlets: drinkable purified water and ultrapure water (resistivity ≥15 MΩ·
Funktionsweise
Tap water → precision sediment + activated carbon pre-treatment removes particles, chlorine, organics → booster pump (0-0.9 MPa) → reverse osmosis (RO) demineralizes water (pressure drives H2O through membrane while ions/organics/colloids are rejected and discharged as concentrate) → polishing mixed-bed (UK Purolite resin) raises resistivity to ≥15 MΩ·cm → 0.2μm terminal microfilter removes residual particles, pyrogens, and bacteria. Anti-scaling RO program periodically flushes the membrane.
Technische Daten
Technische Daten
Abmessungen
Werkstoffe
Wesentliche Merkmale
Anwendungen
Branchen
Komponentenhersteller
Konformität
Reinigung und Wartung
Periodic consumable replacement: 1μm PP cartridge every 1-3 months; activated carbon every 3-6 months; ultrapure cartridges every ~1000-2000L; terminal microfilter every 3-12 months; RO membrane every 1-2 years. If idle >1 month, soak in disinfectant or run periodically.