The SED-CLZ1-K is a premium Class A laminar flow hood with double-layer negative-pressure construction, side-return airflow, EBM imported fan, H14 HEPA, and Dwyer differential-pressure monitoring. Wind speed 0.45 m/s ±20

Grade A (Chinese GMP / EU Annex 1)ISO 14644 Class 5 / FED 209E Class 100H14 HEPA per GB/T 13554
Délai de livraison
Sur devis
Cadence
Contactez-nous

Fonctionnement

The EBM fan draws air through a primary filter into the H14 liquid-trough HEPA filter; the filtered clean air passes through the static pressure box and flow-equalizing membrane for multi-stage pressure equalisation, forming a unidirectional vertical laminar flow at 0.45 m/s ±20% to the work area. The double-layer negative-pressure cavity structure ensures zero leakage; the Dwyer differential-pressure gauge monitors HEPA resistance in real time. PAO/DOP injection and detection ports allow on-site HEPA integrity testing.

Processus de fonctionnement
1The EBM fan draws air through a primary filter into the H14 liquid-trough HEPA filter; the filtered clean air passes through the static pressure box and flow-equalizing membrane for multi-stage pressure equalisation, forming a unidirectional vertical laminar flow at 0.45 m/s ±20% to the work area.
2The double-layer negative-pressure cavity structure ensures zero leakage; the Dwyer differential-pressure gauge monitors HEPA resistance in real time.
3PAO/DOP injection and detection ports allow on-site HEPA integrity testing.

Caractéristiques techniques

Caractéristiques de la variante : SED-CLZ1-K

Caractéristiques

Modèle
SED-CLZ1-K
Tension
220V 50Hz 1Φ
Puissance
0.75 kW

Matériaux

Matériaux en contact
SUS304 acier inoxydable; PAO/DOP test ports SUS304
Matériaux du bâti
SUS304 stainless steel one-piece bent sheet metal

Caractéristiques principales

Imported EBM centrifugal fan (355/310) with large air-volume margin
Double-layer negative-pressure structure — no leakage risk
Multi-stage pressure equalisation with flow-equalising membrane
Wind speed automatically constant or step-adjustable
TKSAGE H14 liquid-trough HEPA, 99.995% @0.3 μm
PAO/DOP injection and detection ports (SUS304 6-point clamps)
SUS304 one-piece bent sheet-metal cabinet, easy to clean and disinfect
Static cleanliness Grade A
Wind speed 0.45 m/s ±20%
Noise ≤65 dB; full-machine vibration amplitude ≤3 μm
Side-return airflow design
0.75 kW 220V VFD-driven fan
Single-unit or multi-unit belt-zone configuration
EBM imported fan + double-layer negative pressure = leak-free Grade A protection
Field-validated PAO/DOP test ports built-in for HEPA challenge testing
Modular — multiple units can be joined to form a belt clean zone over a production line

Applications

Aseptic filling and dispensing protection (Grade A zone)
Sterile sampling, weighing, and assembly
Hospital pharmacy IV admixture
Pharmaceutical packaging line protection

Secteurs

Pharmaceutical (GMP)
Hospital pharmacy
Biotechnology
Cosmetics aseptic manufacturing
Electronics and semiconductor

Marques des composants

Centrifugal fan
EBM (355/310)
VFD inverter
Three Thousand
Differential pressure gauge
Dwyer
Primary filter
TKSAGE nylon-mesh aluminium-frame
HEPA H14 filter
TKSAGE liquid-trough aluminium-frame
LED lighting
HLG T5
PAO/DOP test ports
SUS304 6-point clamp
SED
Échangez avec SED Sales Team
Spécialistes des équipements · contact sur WhatsApp
Consultez les horaires et la disponibilité de l’équipe

Conformité

Grade A (Chinese GMP / EU Annex 1)
ISO 14644 Class 5 / FED 209E Class 100
H14 HEPA per GB/T 13554

Documentation

Fiche technique (PDF)
Aperçu de l’équipement, caractéristiques et dimensions · Document d’origine (anglais)
Demander au service commercial
Manuel opérateur
Installation, fonctionnement et dépannage · Document d’origine (anglais)
Demander au service commercial
Fiche comparative (Standard et Économique)
Comparaison des configurations · Document d’origine (anglais)
Demander au service commercial
Protocoles IQ / OQ / PQ
Documentation de validation pour la conformité GMP · Document d’origine (anglais)
Demander au service commercial
Protocoles FAT / SAT
Essais de réception en usine et sur site · Document d’origine (anglais)
Demander au service commercial
Plans CAD (STEP / IGES / DWG)
Plans d’ingénierie pour l’implantation · Document d’origine (anglais)
Demander au service commercial
Certificats d’étalonnage
Niveau sonore, résistance de terre, rigidité diélectrique et rugosité · Document d’origine (anglais)
Demander au service commercial
DQ (qualification de la conception)
Protocole complet de qualification de la conception · Document d’origine (anglais)
Demander au service commercial
Certificats matières
Composition de l’acier inoxydable 316L · Document d’origine (anglais)
Demander au service commercial
SED
WhatsApp · Ventes SED
Contact commercial · WhatsApp
→