SED-CDC-J — Laminar Flow Pass-Through Box
La description complète de ce modèle est encore en cours de traduction. Consultez les détails en anglais ou demandez des informations à notre équipe. English
The SED-CDC-J is a laminar-flow pass-through box for cleanroom material transfer. The chamber delivers ISO Class 5/US Fed Std 209E Class 100 cleanliness via primary + HEPA two-stage filtration (99.99% @ 0.3 μm). The full
Fonctionnement
The operator places material in the pass-through chamber and closes the door. The double-door electronic interlock prevents both sides from opening simultaneously. Infrared sensing triggers the fan, which draws air through primary and HEPA two-stage filtration to form a laminar flow at 0.45 m/s ±20% inside the chamber. The optional UV germicidal lamp can be activated separately. The Dwyer differential pressure gauge continuously monitors HEPA filter resistance to ensure filtration performance.