SED-0.6CM-J — 0.6 m³, Double-Door Motorized
La description complète de ce modèle est encore en cours de traduction. Consultez les détails en anglais ou demandez des informations à notre équipe. English
SED-0.6CM-J 0.6 m³ horizontal pure-steam sterilizer (autoclave), designed for biopharmaceutical sterile-product manufacturing requiring zero-pyrogen sterilization media. Inner chamber 1,050 × 680 × 850 mm, 6 mm 316L sani
Fonctionnement
Pure-steam pulsating-vacuum sterilization cycle. (1) Load onto 304 SS cart and roll into 316L chamber; both motorized doors close and interlock. (2) Preheat: steam to jacket warms chamber and load to minimize condensate. (3) Pulsating vacuum: water-ring pump alternates vacuum and steam injection to expel residual cold air and ensure uniform penetration of pure steam. (4) Heat-up: chamber temperature rises to 134°C / 0.22 MPa. (5) Sterilization: PLC holds setpoint for programmed time; F0 calculated in real time. (6) Drying: vacuum extraction plus jacket heat dries the load; make-up air enters via 0.22 μm sterile filter. (7) End: alarm; door unlocks; load removed.
Caractéristiques techniques
Caractéristiques
Dimensions
Matériaux
Caractéristiques principales
Applications
Secteurs
Marques des composants
Conformité
Nettoyage et entretien
Mirror-polished 316L chamber — clean with WFI/purified water and compatible alcohol per pharma SOP. Inspect medical silicone door seal each shift; replace per maintenance schedule. Verify 0.22 μm sterile air filter integrity periodically.