Sala de exposición en Ontario, California · Consulte el soporte para su destino
SED-15RXG-C
Consultar
SED-15RXG-C
SED · SED-15RXG-C · Destacado

SED-15RXG-C — Hot Air Circulating Drying Oven

La descripción completa de este modelo aún se está traduciendo. Consulte los detalles en inglés o solicite información a nuestro equipo. English

SED-15RXG-C hot-air circulating drying oven (electrothermal lab/process oven). Working temperature room temp +10 ℃ to 200 ℃; control accuracy ±1 ℃; temperature uniformity ≤±2.5%. Heating power 3.64 kW (two banks of SHEAT

Plazo de entrega
Consultar
Capacidad de producción
Contáctenos

Funcionamiento

Side-mounted blower draws fresh ambient air into the duct system in the bottom/left/right cavities; air passes the U-shaped sheathed heater banks where it is heated, then exits the duct horizontally into the chamber for heat exchange with product on 6 perforated 316-SS trays. Cooled-down air is re-drawn into the duct, mixed and reheated. PID digital controller and K-type SS sensor maintain ±1 ℃ accuracy and ≤±2.5% temperature uniformity.

Proceso de funcionamiento
1Side-mounted blower draws fresh ambient air into the duct system in the bottom/left/right cavities; air passes the U-shaped sheathed heater banks where it is heated, then exits the duct horizontally into the chamber for heat exchange with product on 6 perforated 316-SS trays.
2Cooled-down air is re-drawn into the duct, mixed and reheated.
3PID digital controller and K-type SS sensor maintain ±1 ℃ accuracy and ≤±2.5% temperature uniformity.

Especificaciones técnicas

Especificaciones de la variante: SED-15RXG-C

Especificaciones

Modelo
SED-15RXG-C
Tensión
AC 230V/50Hz
Suministros
AC 230V/50Hz single-phase electrical supply (~3.64 kW peak); flat installation; well-ventilated and grounded location
Detalles de conformidad
Multi-level safety: leakage protection, short-circuit, timer, motor overload, over-temperature.
Potencia
3.64 kW

Dimensiones

Longitud
750 mm
Anchura
1200 mm
Altura
1260 mm
Superficie de instalación
750 × 1200 × 1260 mm (including left control box and top motor)

Materiales

Materiales en contacto
Stainless steel 316 (trays); Stainless steel 304 (inner chamber wall)
Materiales del cuerpo
1.20 mm 304 brushed SS (inner); 1.20 mm cold-rolled steel painted (outer); 4# angle-iron framing (5# base); 100 mm 100K aluminum-silicate fiber insulation; aluminum-silicate door seal

Características principales

Working temperature: room +10 ℃ to 200 ℃
Control accuracy ±1 ℃; uniformity ≤±2.5%
Heating power 3.64 kW (2 banks of sheathed far-IR U-tubes, ~50-60 k h life)
Ramp 7-10 ℃/min
Working chamber 500×600×750 mm; 6 shelves × 316-SS trays (400×600×45 mm)
Forced horizontal air circulation (side fan, 70×210 mm alloy impeller)
Inner 1.20 mm 304 brushed SS; outer 1.20 mm painted cold-rolled steel
100 mm 100K aluminum-silicate insulation
100 mm-thick single door with high-temp aluminum-silicate seal
Yuyao Cheng Neng PID controller; K-type SS sensor
SSR (silent) contactor; Chint switchgear / over-temp meter
Protections: leakage, short-circuit, timer, motor overload, over-temperature
Single-phase AC 230V/50Hz supply (lab-friendly)
316-SS trays included (pharma-grade contact surfaces)
Long-life sheathed far-IR heaters (50-60 k h)
Custom paint color per customer specification
1-year whole-machine warranty

Aplicaciones

Drying and curing of pharmaceutical materials and granules
Lab and small-batch process drying
Heat treatment of plastic, rubber, and electronic components

Sectores

Pharmaceutical (lab / R&D / small batch)
Chemical / fine chemical
Electronics
Universities & research institutes

Marcas de componentes

Temperature controller: Yuyao Cheng Neng PID digital (1 ℃ resolution)
Temperature sensor: K-type stainless steel (1 unit)
Solid-state contactor: SSR (silent operation)
Over-temperature meter: Chint
Switches & indicators: Chint
General electrical components: Chint
Heater: SHEATHED HEATER (far-IR U-shaped SS tube, 2 banks)
SED
Hable con SED Sales Team
Especialistas en equipos · contacto por WhatsApp
Consulte el horario y la disponibilidad del equipo

Limpieza y mantenimiento

Operate at 15-35 ℃ ambient. Avoid strong vibration, EMI, dust, corrosive substances, direct sunlight, direct heat radiation, and direct air drafts on the cabinet.

Documentación

Ficha técnica (PDF)
Resumen del equipo, especificaciones y dimensiones · Documento original (inglés)
Solicitar al equipo comercial
Manual del operador
Instalación, funcionamiento y diagnóstico · Documento original (inglés)
Solicitar al equipo comercial
Ficha comparativa (Estándar y Económica)
Comparación de configuraciones · Documento original (inglés)
Solicitar al equipo comercial
Protocolos IQ / OQ / PQ
Documentación de validación para conformidad GMP · Documento original (inglés)
Solicitar al equipo comercial
Protocolos FAT / SAT
Pruebas de aceptación en fábrica y en sitio · Documento original (inglés)
Solicitar al equipo comercial
Planos CAD (STEP / IGES / DWG)
Planos de ingeniería para la instalación · Documento original (inglés)
Solicitar al equipo comercial
Certificados de calibración
Nivel sonoro, resistencia de tierra, rigidez dieléctrica y rugosidad · Documento original (inglés)
Solicitar al equipo comercial
DQ (cualificación del diseño)
Protocolo completo de cualificación del diseño · Documento original (inglés)
Solicitar al equipo comercial
Certificados de materiales
Composición del acero inoxidable 316L · Documento original (inglés)
Solicitar al equipo comercial
SED
WhatsApp · Ventas SED
Contacto comercial · WhatsApp
→