Ausstellung in Ontario, Kalifornien · Support am Zielort anfragen
Startseite / Maschinen / Arzneimittel-Stabilitätsprüfkammer / Pharmaceutical Stability Test Chamber (500 L) — YPX-S Series
SED-500YPX-S
Auf Anfrage
SED-500YPX-S YPX-S Series
SED · SED-500YPX-S · 1 verkaufte Einheiten

SED-500YPX-S — YPX-S Series

Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English

SED-500YPX-S Pharmaceutical Stability Test Chamber (500 L). Constant-temperature constant-humidity chamber for ICH-Q1A(R2)/Q1B and 2020 Chinese Pharmacopoeia stability testing — long-term, accelerated, intermediate, phot

Compliant with JJF1101-2019 environmental test equipment temperature/humidity calibration specificationCompliant with 2020 China Pharmacopoeia drug stability testing guidelinesCompliant with ICH 2003 Q1A(R2) and Q1B
Variante wählen (4)
Lieferzeit
Auf Anfrage
Durchsatz
Kontaktieren Sie uns

Funktionsweise

Forced air circulation balanced temperature/humidity control. Centrifugal fan circulates air through duct containing heater, evaporator, humidifier and dehumidifier sections; conditioned air enters chamber uniformly. Heating: nickel-chromium electric element with PID control via SSR (solid-state relay, zero-cross firing). Refrigeration: imported fully-sealed compressor with R134a (DuPont). Humidification: built-in electric steam (water from internal pump, closed-loop). Dehumidification: refrigeration coil. Temperature sensor: PT100 Class A (Yifan/Taiwan); humidity sensor: Vaisala Y60 (Finland).

Prozessbeschreibung
1Forced air circulation balanced temperature/humidity control.
2Centrifugal fan circulates air through duct containing heater, evaporator, humidifier and dehumidifier sections; conditioned air enters chamber uniformly.
3Heating: nickel-chromium electric element with PID control via SSR (solid-state relay, zero-cross firing).
4Refrigeration: imported fully-sealed compressor with R134a (DuPont).
5Humidification: built-in electric steam (water from internal pump, closed-loop).
6Dehumidification: refrigeration coil.
7Temperature sensor: PT100 Class A (Yifan/Taiwan); humidity sensor: Vaisala Y60 (Finland).

Technische Daten

Technische Daten für Variante: SED-150YPX-S 150 L

Technische Daten

Modell
SED-500YPX-S
Spannung
AC 220V / 50Hz (single phase typical for this size)
Versorgung
Electrical: AC 220 V/50 Hz, 3 kW rated maximum. Water: deionized water supply for internal humidification reservoir (auto-fed via small internal pump). Drain not strictly required — closed water loop.
Konformitätsangaben
12-month warranty after customer acceptance with 24-hour response, 2-day on-site arrival, 3-day fix SLA. Lifetime maintenance after warranty (parts at cost). Full 3Q validation file (DQ/IQ/OQ + FAT/SAT/PQ) included on USB. Compliant with Chinese GMP drug manufacture quality management regulations.
Leistung
3 kW

Abmessungen

Länge
1020 mm
Breite
1100 mm
Höhe
1720 mm
Aufstellfläche
Floor-standing on 4 lockable casters. External dimensions 1020×1100×1720 mm (W×D×H). Allow rear and side clearance for ventilation and door swing.

Werkstoffe

Kontaktwerkstoffe
Mirror-finish stainless-steel inner liner
Gehäusewerkstoffe
Cold-rolled steel outer cabinet, phosphated and electrostatically powder-coated; polyurethane foam insulation (water-resistant, environmentally friendly); high-temperature silicone-rubber door gasket

Wesentliche Merkmale

Volume: 500 L (internal W×D×H = 800×700×900 mm)
Temperature range 10–65 °C; control fluctuation ±0.5 °C; deviation ±2.0 °C; uniformity 2.0 °C
Humidity range 20–98% RH; uniformity/deviation 5.0% RH
Optional photostability lighting 0–10,000 lx (deviation ±500 lx); near-UV 300–400 nm
Touchscreen PLC controller; resolution 0.1 °C / 0.1% RH; PID self-tuning
Three-tier permission (admin / operator / visitor) with audit-trail logging
Built-in 4 GB storage (~1 year data); USB export (~5 years)
Dot-matrix printer for time/temperature/RH/lux records
Vaisala (Finland) humidity sensor; Class-A PT100 temperature sensor
Imported fully-sealed compressor (R134a refrigerant)
Two adjustable shelves (≥10 kg load each)
Single door with lock + tempered-glass observation window + Phi 50 mm test port on left side
Four casters with brakes — movable
Independent over-temperature protection (separate sensor cuts heater)
Audible/visual alarm + SMS/app push for over-temp/RH, low water, compressor overload, power-loss
ICH/Pharmacopoeia-compliant stability chamber — required for GMP drug registration
Three-tier permission + audit trail required by data-integrity (ALCOA+) audits
Vaisala humidity sensor — gold standard in pharmaceutical stability testing
Full validation document set provided (DQ, IQ, OQ, PQ, FAT, SAT) on USB
12-month warranty with 24-hour response, 2-day on-site, 3-day fix commitment; lifetime maintenance after warranty
Optional SMS/mobile-app alarm for unattended overnight runs

Anwendungen

ICH long-term stability test (25 °C/60% RH; 30 °C/65% RH; 5 °C; -20 °C)
Accelerated stability test (40 °C/75% RH for 6 months)
Intermediate stability test (30 °C/65% RH for 12 months)
Influencing-factor tests (high temp 40/50/60 °C, high RH 25 °C/90% RH, photostability 4500 lx ±500 lx with near-UV)

Branchen

Pharmaceutical manufacturing and R&D
Biotechnology
Medical and life sciences research

Komponentenhersteller

Compressor: imported fully-sealed with DuPont R134a refrigerant
Temperature sensor: Yifan (Taiwan) Class-A PT100 armored sensor
Humidity sensor: Vaisala Y60 (Finland)
Controller: touchscreen PLC with PID self-tuning
Contactor / breaker / relays: Schneider
SSR: solid-state relay (zero-cross power control)
Printer: dot-matrix microprinter
Casters: four movable casters with brakes
SED
Sprechen Sie mit SED Sales Team
Maschinenspezialisten · Kontakt per WhatsApp
Geschäftszeiten und Verfügbarkeit beim Team erfragen

Konformität

Compliant with JJF1101-2019 environmental test equipment temperature/humidity calibration specification
Compliant with 2020 China Pharmacopoeia drug stability testing guidelines
Compliant with ICH 2003 Q1A(R2) and Q1B

Reinigung und Wartung

Mirror stainless-steel interior wipes clean. High-temperature silicone door gasket maintenance per schedule. Refill humidifier water reservoir with deionized (DI) water periodically via internal pump.

Dokumentation

Datenblatt (PDF)
Überblick, technische Daten und Abmessungen · Originaldokument (Englisch)
Beim Vertrieb anfragen
Bedienhandbuch
Einrichtung, Betrieb und Fehlersuche · Originaldokument (Englisch)
Beim Vertrieb anfragen
Vergleichsdatenblatt (Standard und Economy)
Vergleich der Konfigurationen · Originaldokument (Englisch)
Beim Vertrieb anfragen
IQ / OQ / PQ-Protokolle
Validierungsunterlagen zur GMP-Konformität · Originaldokument (Englisch)
Beim Vertrieb anfragen
FAT / SAT-Protokolle
Abnahmeprüfungen im Werk und vor Ort · Originaldokument (Englisch)
Beim Vertrieb anfragen
CAD-Zeichnungen (STEP / IGES / DWG)
Konstruktionszeichnungen für die Aufstellung · Originaldokument (Englisch)
Beim Vertrieb anfragen
Kalibrierzertifikate
Schallpegel, Erdungswiderstand, Spannungsfestigkeit und Rauheit · Originaldokument (Englisch)
Beim Vertrieb anfragen
DQ (Designqualifizierung)
Vollständiges Designqualifizierungsprotokoll · Originaldokument (Englisch)
Beim Vertrieb anfragen
Werkstoffzeugnisse
Zusammensetzung des Edelstahls 316L · Originaldokument (Englisch)
Beim Vertrieb anfragen
SED
WhatsApp · SED Vertrieb
Vertriebskontakt · WhatsApp
→