SED-1MM-J
Auf Anfrage
SED-1MM-J
SED · SED-1MM-J · Hervorgehoben
SED-1MM-J — Autoklav mit pulsierendem Vakuum
Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English
Pulsating-vacuum autoclave (sterilizer) for hospitals, pharma factories, health and research units. Sterilizes clothing, dressings, metal instruments, culture media and glassware using saturated steam. Horizontal double-
Variante wählen (13)
Lieferzeit
Auf Anfrage
Durchsatz
Kontaktieren Sie uns
Funktionsweise
Process flow: auto-prep → pulsating vacuum → heat-up → sterilize → dry → end. Pulsating vacuum removes cold air pockets; saturated steam holds at 134 °C / 0.22 MPa; post-sterilization vacuum dehumidification with jacket drying yields dry loads.
Prozessbeschreibung
1Process flow: auto-prep → pulsating vacuum → heat-up → sterilize → dry → end.
2Pulsating vacuum removes cold air pockets; saturated steam holds at 134 °C / 0.22 MPa; post-sterilization vacuum dehumidification with jacket drying yields dry loads.
Technische Daten
Technische Daten für Variante: SED-0 .24MM-J (700×600×600 chamber)
Technische Daten
Modell
SED-1MM-J
Spannung
AC380V/50Hz/3-phase 5-wire AC380V/50Hz
Versorgung
Steam 0.4–0.6 MPa; water 0.1–0.3 MPa; compressed air 0.5–0.7 MPa; power 3-phase 5-wire AC380V/50Hz 4.5 kW
Leistung
4.5 kW
Abmessungen
Länge
1530 mm
Breite
1480 mm
Höhe
2000 mm
Gewicht
1500 kg
Werkstoffe
Kontaktwerkstoffe
6 mm SUS304 inner chamber; SUS304 sanitary piping; SUS316 sterilization carts
Gehäusewerkstoffe
6 mm Q345R pressure-vessel plate jacket; δ=1.2 mm SUS304 brushed-stainless decorative cover; 50 mm aluminium-silicate insulation
Wesentliche Merkmale
Horizontal double-layer: SUS304 6 mm inner chamber + Q345R 6 mm jacket
Pulsating-vacuum cold-air removal for uniform temperature
Post-sterilization vacuum dehumidification + jacket drying
Working pressure 0.22 MPa / design pressure 0.245 MPa
Working temperature 134 °C / design temperature 139 °C
Thermal uniformity ≤±1 °C
Ultimate vacuum -0.095 MPa
Standard GMP validation interface
Siemens S7-200 PLC + touchscreen HMI + micro printer
Electronic signature, admin rights, dual temperature probe recording
Real-time temperature/pressure/time/F0 reports and trend charts
US ASCO original pneumatic valves (4 M-cycle life)
Water-ring vacuum pump: low noise, high vacuum
0.22 µm sterile air filter
Pneumatic valve + steam trap eliminates check-valve back-leakage
Double automatic doors with safety interlock
Medical silicone seal with compressed-air auto-sealing
X-ray flaw detection on welds
SUS316 sterilization car and trolley supplied for ≥1.0 m³ units
Transport rating -40 °C to +50 °C, RH 10–93%
Chamber volume 1.0 m³; overall 1530×1480×2000 mm; weight 1500 kg; power 4.5 kW
GMP and GB150-2011 pressure-vessel compliance
Premium Siemens PLC, ASCO valves, 0.22 µm sterile filter
≤±1 °C uniformity for repeatable sterilization cycles
Full data records with e-signature for archiving
Double-door interlock separating clean/non-clean sides
Anwendungen
Saturated-steam sterilization of clothing, dressings, metal instruments
Culture-media and glassware sterilization
PIX packaging and inactivation tank systems (optional)
B&D test programs
Branchen
Hospital
Pharma
Health
Scientific research
Komponentenhersteller
PLC S7-200
Siemens
Pneumatic angle-seat valve
ASCO
Vacuum pump
Water-ring type (Nash/Nassim style)
Sealing ring
Medical silicone
SED
Sprechen Sie mit SED Sales Team
Maschinenspezialisten · Kontakt per WhatsApp
Geschäftszeiten und Verfügbarkeit beim Team erfragen
Konformität
GMP design and validation interface
GB150-2011 steel pressure vessel
X-ray flaw inspection certificate
Dokumentation
Datenblatt (PDF)
Überblick, technische Daten und Abmessungen · Originaldokument (Englisch)
Beim Vertrieb anfragen
Bedienhandbuch
Einrichtung, Betrieb und Fehlersuche · Originaldokument (Englisch)
Beim Vertrieb anfragen
Vergleichsdatenblatt (Standard und Economy)
Vergleich der Konfigurationen · Originaldokument (Englisch)
Beim Vertrieb anfragen
IQ / OQ / PQ-Protokolle
Validierungsunterlagen zur GMP-Konformität · Originaldokument (Englisch)
Beim Vertrieb anfragen
FAT / SAT-Protokolle
Abnahmeprüfungen im Werk und vor Ort · Originaldokument (Englisch)
Beim Vertrieb anfragen
CAD-Zeichnungen (STEP / IGES / DWG)
Konstruktionszeichnungen für die Aufstellung · Originaldokument (Englisch)
Beim Vertrieb anfragen
Kalibrierzertifikate
Schallpegel, Erdungswiderstand, Spannungsfestigkeit und Rauheit · Originaldokument (Englisch)
Beim Vertrieb anfragen
DQ (Designqualifizierung)
Vollständiges Designqualifizierungsprotokoll · Originaldokument (Englisch)
Beim Vertrieb anfragen
Werkstoffzeugnisse
Zusammensetzung des Edelstahls 316L · Originaldokument (Englisch)