SED-1.0MM1-A (2) — Autoklav mit pulsierendem Vakuum
Die vollständige Beschreibung dieses Modells wird noch übersetzt. Lesen Sie die englischen Details oder fragen Sie unser Team nach weiteren Informationen. English
Pure-steam pulsed-vacuum sterilizer model SED-1.0MM1-A, 1000 L chamber, double-door design (loading door in Class B zone, unloading door in Class A zone, mechanically interlocked so both doors cannot open simultaneously)
Funktionsweise
Double-jacketed vessel: pure steam in inner chamber for sterilization, raw boiler steam in jacket for vacuum-exhaust and drying. PLC + HMI controls run pre-set cycles: Bowie & Dick test (134°C/15 min), vacuum-leak test, porous-load (cloth/tools) sterilization with initial pulsed vacuum + sterilization + vacuum drying, and culture-medium cycles. Three PT100 temperature probes (2 movable inside chamber + 1 fixed at drain) ensure temperature uniformity ≤±1°C. Safety device returns chamber to atmospheric pressure on controller failure to allow loading-door opening.
Technische Daten
Technische Daten
Abmessungen
Werkstoffe
Wesentliche Merkmale
Anwendungen
Branchen
Komponentenhersteller
Konformität
Reinigung und Wartung
Calorie-hoop interlinkage fast joints (sanitary clamps) on product-contact piping for easy disassembly, cleaning, and parts replacement. Long-flow pure steam and air-fill pipework design ensures clean energy delivery.